· Balloon Catheters
Mozec PTA Balloon Dilatation Catheter 0.014"
| Technical Blueprint Specification | Clinical Value |
|---|---|
Usable Shaft Length | for (OTW 0.014") - 120cm ± 2cm & 150cm ± 2cm, for (OTW 0.035") - 80cm ± 2cm & 135cm ± 2cm |
Shaft Diameter | for (OTW 0.014") - Distal Shaft 0.95mm (2.9Fr) & Proximal Shaft 1.32mm (4.0Fr), for (OTW 0.035") - Balloon diameter 3 to 10mm: 5.0 Fr(1.72 ± 0.025mm), Balloon diameter 12 to 14mm : 6.0 Fr(1.98 ± 0.025mm) |
Max. Guidewire Compatibility | 0.014" / 0.035" |
Balloon RO Markers | 2 Platinum-Iridium |
Tip Length | 5.0 ± 1mm |
Nominal Pressure (NP) | 6-7 atm. depending upon diameter & length of balloon (Refer labels/IFU for more details) |
Rated Burst Pressure (RBP) | for (OTW 0.014"): 12-16 atm. depending upon diameter & length of balloon (Refer labels/IFU for more details), for (OTW 0.035") : 11-14 atm. depending upon diameter & length of balloon (Refer labels/IFU for more details) |
Catheter System
Over the wire (OTW)
Balloon Material
PEBA
Distal Shaft Coating
Biocompatible, Hydrophilic & Lubricious Coating
Primary Material
PEBA
Clinical Sizing Precision
Over the wire balloon dilatation catheter for percutaneous transluminal angioplasty in carotid, iliac, femoral, popliteal, infra-popliteal and renal arteries. Features assured trackability, responsive pushability, and superior deliverability with hydrophilic coating.
This product is manufactured by Meril Life Sciences Pvt. Ltd. and is distributed by Agile Healthcare across all districts of Telangana.
Yes, all Meril Life Sciences products distributed by Agile Healthcare, including the Mozec PTA Balloon Dilatation Catheter 0.014", strictly adhere to CDSCO guidelines and ISO 13485:2016 quality standards.
The Mozec PTA Balloon Dilatation Catheter 0.014" is constructed using PEBA, ensuring high biocompatibility and durability.
Balloon Catheters
Balloon Catheters
Balloon Catheters
Balloon Catheters